07-21-2005 06:52 PM
07-22-2005 12:27 AM
07-22-2005 03:42 AM
Is this drift permanent? Say, you have balanced your bridge and you continiusly have to decrease your balance dial for zero?
What type of strain sensor are you using?
07-22-2005 07:36 AM
07-22-2005 11:00 AM
First... thanks for all the suggestions... to answer in the order which they were recieved...
1) Hardware Configuration:To the best of my knowledge, it is correct. I am using a DAQCard-6062E for data acquiusition, running the incoming data through a 5B01 backplane with 5B38 isolation/ amplification modules to better see the signal and prevent frying the card in the event something stupid is done. I do not believe this is an issue due since I have other channels sampled which stay constant over time which should drift if it was a configuration problem.
2) Additional Information: Currently using a 350 ohm strain sensor from Micro-Measurements across a wheatstone excited with 10Volts in a quarter bridge configuration. Gage
3) Does the same drift occur via voltmeter... Good Question... didn't even thnk to check that part directly, as LV has treated me well otherwise. Unfortunately, Yes, the VM output also drifts over time.
07-22-2005 11:20 AM
Sorry, sometimes my premature posting problem gets out of control...
2) continued...
Gage information: (Micro-Measurements) CEA-06-125UW-350
Bridge information: (Micro-Measurements) MR1-350-130
Isolation Module: (Analog Devices) 5B38 Isolated Strain Gage Input (wide bandwidth)
Bridge Balancing: This I am performing in LV using a shift register to return the previous loop value to the 'Vinit' terminal of the strain gage conditioning vi. Thus the bridge is balanced using an offset in LV, not on the bridge itself. (I'm performing this project for someone that knows little about such things... and would rather have him press a button on the front panel to initialize the bridge... than to fiddle with a balancing resistor.... though if I can map the drift with no load and reliably replicate it, I can set up a program to scale the data once it is catalogued. Would be a messy way of doing things... but might work. (Brainstorming with the keyboard.))
Ok, I think that is about all there is for now... Again, Many thanks for all suggestions.
Paul
--- LV 6.0.2
07-22-2005 03:02 PM
07-22-2005 03:45 PM