Hello JohnnyF,
Thank you for posting to the discussion forum. It looks like you are trying to use the NI-9233 module. Are you using the USB-9162 carrier or are you using the cDAQ Chassis?
If you are using the cDAQ Chassis, reference the following user guide to find out how to setup the hardware and the software installation.
NI cDAQ-9172 User Guide and Specifications
If you are using the USB-9162 carrier with the NI-9233 module, you should reference the following documentation:
USB-9233 User Guide and Specifications
NI-DAQmx for USB Devices Getting Started Guide
After getting the hardware installed and Measurement and Automation Explorer up and running, run the self test. After passing the self-test you can then open a test panel to measure an analog input signal. You can also configure a task to read information from your load cell.
For some load cells you can reference the following KnowledgeBase:
How do I Configure a Load Cell in MAX?
In this case though, you are using the 9233 which won't allow you do a custom voltage with excitation with DAQmx. I would recommend creating a Voltage Task to read the voltage and then use a custom scaling to scale the values. You need to reference your certificate of calibration for your load cell to determine the corresponding scaling values. You may need to contact the vendor of the dynamic load cell to get this information.
Regards,
Sandra T.
Applications Engineer | National Instruments