07-15-2009 09:42 PM
Does anyone know how to initialize PXI DMM without opening DMM soft front panel when I start PC?
Thanks,
Adam
Solved! Go to Solution.
07-16-2009 09:05 AM
07-16-2009 09:53 PM
Hi, Adam
Please add the ini files for initializing the instrument when OS starting up. While OS loading the ini files, here sometimes comes an error message. It occurs to OS loading the ini files before the environment being ready. Therefore, adding a waiting time, ex. 10s, can solve this error.
Thanks.
NITW