11-17-2000 08:33 AM
11-17-2000 06:12 PM
08-03-2001 10:35 AM
04-16-2007 11:44 AM
Hi, I have a similar problem with calibration files for HP8753D. Actually, I try to save the calibration from my Analyzer to my PC and it doesn't work. I mean, it is saved but I can not read the file.
Could you explain me, why?
04-17-2007 07:01 PM - edited 04-17-2007 07:01 PM
Message Edited by Adam W on 04-17-2007 07:01 PM
07-04-2008 05:01 AM
I have recently run into the same problem, and may have a solution that I am yet to prove, although it is a possible problem in the driver anyway
When loading calibration data into the network analyser the transfer would sometimes fail. I have not had time to re-try this out although i think i have found the problem. In the Cal file they use the POIN; command say for example POIN201;
From the Manual :
"This command should be followed by a wait command equal to at least 2 sweep times"
The driver doesnt implement this and could be the source of the timing issue. Will update next time I have chance to get on the analyser.
Craig