12-02-2006 07:37 AM
12-02-2006 09:03 AM
Thanks for all time you are spending. Some other info, maybe helpful. The multimeter works also with a software called Mas-View. I monitored the communication with this software too. please fn in the file the output of portmon. The software doesn't use Visa driver I suppose. So that is seems to me that the communicatio is cleaner.
Both Matlab and Mas-View seem to get the answer back from the instrument in a different way. Labview cannot.
Regards
Gianmarco
12-02-2006 02:53 PM
12-03-2006 05:29 AM
12-03-2006 05:57 AM
12-06-2006 01:52 PM
Hi guys,
after several trials and good suggestions from NI support team (Italy) I finally solved my problem. Please find enclosed .vis for a working communication to:Digital Multimeter DVM345DI ex Velleman (alias MAS34-xx ex Mastech).
For the beginners like me, I suggest to measure the voltage of each pins in the serial port during a working communication when they want to know what is going on. So I did with the multimeter itself. I understood that the working programs (Mas-view and Matlab) setted the RTS to null instead Labview put it to high. I suppose this is not a very common serial transmission procedure but that is.Thanks to people that helped me.
Gianmarco
12-07-2006 04:58 AM
12-17-2006 11:40 PM
12-18-2006 01:21 PM
12-18-2006 05:38 PM