Hi,
I'm working in labview to create a PID control loop with a flow meter and pressure sensor for pressure control. However, I cannot get reliable results from the pressure transducer using a PCI-6024E aquistion card and labview 6.0. My problem is that when i connect the pressure transducer wheatstone bridge across differential inputs to the DAQ card, I get a voltage offset. I have checked using the MAX test panel that my input is connected to the correct channels under a diff input setup, and have even setup a test bridge from resistors but cannot get rid of the offset.
For example when applying 5V across a wheatstone bridge made up of 4 562 ohm resistors, if i connect the DAQ inputs I get a reading of 0.13V. I
have the AIGND and power supply earth connection connected to the base of the bridge so i don't think it is a grounding problem.
If I probe the bridge without the DAQ connected I get a reading of 2.5V from both sides of the bridge. however as soon as i connect the DAQ card, one side of the bridge (the one connected to the higher of the pair of diff input channels) reads 2.37V, creating this offset.
I've tested using other channels on the DAQ so its not that one pair are a little dodge.
I've spent all week frustrated by this problem, any help is much appreciated!
Rob.