04-10-2008 10:25 AM
04-10-2008 10:56 AM
Why wouldn't the sample rate decrease? For one measurement, you send a command to the instrument to take a measurement and then retrieve the data. This takes x amount of time. You then send an almost identical command and read the results. This is going to take x amount of time as well. There may be some slight differences in the time it takes the instrument to perform a voltage measurement and a current measurement but probably not enough to notice.
Whether or not the instrument is capable of doing simultaneous measurements is something I don't know. You will have to read the manual to see whether you can set it up for this. With some instruments, you setup the measurement types and when you ant to perform it, issue an initiate command and do a Fetch for each reading. Certain instruments differentiate between a MEAS?, READ?, and a FETCH? A MEAS? command will usually force the instrument to initiate a new measurement each time and a Fetch? will only tell the instrument to send the current data. I don't have an Amrel any more and I can't download the driver so I'm not sure what the driver is using.