Hi,
I have connected a USB 6008 to my computer but cannot read any output from it using a multimeter when I run my programme.
I've gone into the measurement and automation explorer, gone to the USB6008 device, clicked test panels and selected voltage output. From here I've gave varying output voltages and I can read all of these voltages using a multimeter from the USB 6008.
To check whether there was something wrong with the block diagram etc. that I had setup, I connected the 6008 to another computer which runs labview and used the same programme. It is giving an output fine.
Does anyone have any suggestions as to what may be wrong on the first computer?
Much appreciated,
Paul
Solved! Go to Solution.
11-12-2008 10:32 AM
11-12-2008 10:44 AM
Thanks for reply,
Yes, I've run a self test and it passed. In the Daq assistant menu, on the computer which won't work, the terminal configuration drop down menu will only allow the 'let NI-DAQ choose' whereas on the computer on which it works, it will allow the RSE configuration setting to be selected. Does this give any clue as to what may be wrong?
Paul
11-12-2008 10:46 AM - edited 11-12-2008 10:48 AM
11-12-2008 10:57 AM - edited 11-12-2008 10:58 AM
I am using LAbview 8.0 on both machines.
I'm using Windows XP on both machines also.
11-12-2008
11:04 AM
- last edited on
06-17-2024
02:27 PM
by
Content Cleaner
If I were in your shoes I would download DAQ-mx 8.6.1 (here) on the PC you are having trouble with and give it a go. I don't know why the 6008 would work on one PC and not the other when they both have the same setup. Do you have other devices working on the troubled PC?