Hi Manuel,
I'm guessing you installed the IVI driver for the Tecktronix scope. Based on the error, it is likely that you do not have VISA (or some other dependent software component) installed on your system. When VISA is not present on a system, the IVI engine fails to load. This causes functions that use the IVI engine error -1074135040 (xBFFA0000). You can download NI-VISA from NI's support and downloads page. Check the link below to be sure that you hav ethe correct version.
https://www.ni.com/en/support/downloads/drivers/download.ni-visa.html
Do you have downloaded the instrument driver on this web page https://sine.ni.com/apps/utf8/niid_web_display.model_page?p_model_id=9076 ?
If no, try with this one.
Regards,
Christophe S.
Account Manager East of France І Certified LabVIEW Associate Developer І National Instruments France
12-19-2007 05:03 PM
12-20-2007
01:46 AM
- last edited on
07-13-2026
08:59 AM
by
Content Cleaner
Hello Joao,
And do you have installed the IV compliance Engine ?
Check this link https://forums.ni.com/t5/LabVIEW/Error-1074135040-Revisited/m-p/274216?requireLogin=False, there is a similat post with the same error message.
Regards,
Christophe S.
Account Manager East of France І Certified LabVIEW Associate Developer І National Instruments France