It is not possible to combine different measurement types with DAQ Assistant. However, you can combine different types programmatically. Click
here for information on how to do that.
To change the x-axis scale to relative time, right click the Graph in the front panel. Go to X-scale -> Formatting. Under the �Format and Precision� tab, select �x-axis� from the drop down menu. Then select �Relative Time� from the selection box.
The SCXI-1102 has programmable gain amplifiers that can be programmed for a gain of 1 or 100. If you programmatically set the gain to 1, then the SCXI would no
t amplify the signal.