We have a large number of SCXI systems and they are very good about eliminating noise. In software, I used a 4 Hz filter and take the average of 100 readings each second. We are trying to move to CompactDAQ for several reasons, but it seems as though the voltage readings using a 9205 DSUB are much suceptible to noise.
StDev for SCXI - 0.0089
StDev for cDAQ - 0.0206
The "average" lines in the attached chart are 10 second sliding window averages.
Why can't I use a 4 Hz filter in Compact DAQ? Not supported?
Is this issue because SCXI channels are isolated? (SCXI-1125)
We are using an in house bias resistor board for the Compact DAQ and an SCXI-1328 for SXCI.
Thanks