05-23-2007 03:50 PM
05-25-2007
10:19 AM
- last edited on
03-10-2025
12:59 PM
by
Content Cleaner
Craig,
If you are talking about the PCI-TIO-10, there is a user manual for that. Please click here to access the manual. Also, if you tell me what your application is, I can recommend a newer card for you.
Regards,
05-27-2007 03:23 PM
Thanks, yes the card is a PC-T10-10, we are new to DAQ and didnt know quite what we were looking at but that part number is stamped on it. The software is a custom written package for reading leakage current from high voltage equipment. We cant configure it at all and the OEM is none too helpful. I did put a question into NI's "ask an expert" to get an opinion on wether we change the card and change the application to Dasylab or something so we arent flying blind in a fault. Leakage current from four items of HV plant in a substation (0-20mA) is passed through a resistor, the voltage tapped off and converted V/F to drive a fibre transmitter at (we think about) 300Hz + leakage value and fibered inside to the PC. The fibre receivers (on a seperate interface board before the pc) appear to feed the freq straight into the PC-T10-10 for conversion back to leakage current values (hard to tell without a schematic). Each of the four channels has two relay outputs also, a leakage warning at 15mA and an alarm at 20mA. Have just started fault finding now and were looking for info. It may be that we scrub it all and start again with something a bit more mainstream.
05-27-2007 05:32 PM
05-30-2007 02:06 PM
05-31-2007 06:17 PM
06-04-2007
12:12 AM
- last edited on
03-10-2025
01:01 PM
by
Content Cleaner
Craig,
If you have tried multiple computers and are seeing the same behavior, then a bad card might be the case. There is no built software to check the working of your card. However, if you are looking to upgrade your computer and system, we do have some other newer DIO cards, which I can recommend. The 653X series are cards that are capable of Digital I/O and timing. The 6536/7 are PCI-Express compatible as well making data transfer much faster than the PCI cards. If you could tell me a little more about your entire application I can narrow down my recommendation as well.
You can find more information and specifications on these DIO cards here.
Regards,
05-25-2011 08:08 PM
I'd like to know what is a newer PCI form factor replace for PC-T10-10 181197-01 I/O Card
Thx.
--Fred
05-26-2011
01:51 PM
- last edited on
03-10-2025
01:01 PM
by
Content Cleaner
Hi Fred, this is Paul with Applications Engineering at NI.
In the future, it would probably be better to start a new thread for a question like this.
As for replacing this card, we have options available. If you'd like to share more about your application we can evaluate what card might be the best fit for you. Important things to include would be: the signal types you want to measure (analog, digital, etc), the frequency range you are working in, and how many channels you require.
Regards,