01-23-2024 01:35 AM
Hi Guys,
We are using more than 60 pins from 2 different NI 6571 Card in same pattern, we have faced the delay between Some channels belongs to different NI 6571 card. We suspect the vectors of 2 cards are not started at same point. Is it possible in NI STS chasis with preinstalled NI 6647T?
We have understood that this may get solved using N - ClK syncronization. But we face the below error. Is any possible way to solve this error
we are using NI Digital 19.0 Version , Labview and testand of 2019 version.
Kindly review our issue and solution and help us by confirming the issue and our way of addressing the same in right path.
Thanks ,
Yohes
01-23-2024 08:36 AM
You can let go of all those 6674 TCLK Sync items and just put all the 6571s that are part of those 60 pins into the same Digital Group in PinMap.
TSM 2019 shipped with the feature to have multi-instrument sessions for NI-Digital.