Inconsistent READ output at NI-MAX and Labview for Serial Port Device (mk2000)

Solved!
Go to solution

Sorry.

I will send them in 2018 later today. I can save for previous version.

greetings from the Netherlands
Message 11 of 14
(744 Views)

That will be great Albert. Thank you !!!!

0 Kudos
Message 12 of 14
(740 Views)

The 20018 versions, just saved for previous

greetings from the Netherlands
Message 13 of 14
(735 Views)
Solution
Accepted by topic author kharkhua

Thanks a lot Albert. Great Help !!!

 

Turns out that putting a VISA flush block before read solved the issue of lag between command and output. Also I had to manually add ''0A" end line character in hexa display along with command to avoid the errors. 

 

I am sharing a simple code to write and read command for mk2000.Simple W&R code for mk2000Simple W&R code for mk2000

0 Kudos
Message 14 of 14