If you mean sending trigger signal to another device, you could use niScope Export Signal.vi
10-13-2010 12:57 AM
Can I do like this,
Channel 0: data input
Channel 1: another data input
PFI1: A sine wave input, using the wave cross 0 point to trigger data collection for CH0 and CH1.
10-13-2010 05:04 AM
PFI only supports digital trigger. You need to build comparator circuit with opamp to change from sine wave to 0/5V signal.
10-13-2010 08:23 AM
Is there any detail reference of digital trigger.
Is it triggerred by voltage level or rising edge? Does it has the same accuracy with analog trigger. Thanks.
10-13-2010 08:42 AM
10-13-2010 09:20 PM
Hi KateB,
Thanks for your info about PFI in 5133. Is there any info about general description about Digital Trigger, thanks.