Garbage being displayed

I am not exactly sure what is going on. We have a program that does some serial communication via RS485. The program works fine pre-build but when it is built I am getting odd characters (wing-ding type). It is on a different PC so maybe it is a driver issue with the RS485 card..
 
 
Anyone have an idea of why this would be?
0 Kudos
Message 1 of 5
(3,186 Views)
What do the characters look like? (A screen shot would help)

What are the hex/decimal values that are being returned?

Mike...

Certified Professional Instructor
Certified LabVIEW Architect
LabVIEW Champion

"... after all, He's not a tame lion..."

For help with grief and grieving.
0 Kudos
Message 2 of 5
(3,169 Views)
I've often seen this happen when Baud rates are incorrect.
---------------------
Patrick Allen: FunctionalityUnlimited.ca
Message 3 of 5
(3,148 Views)
Have seen such a behaviour (also in Dev Enviornment) with an instrument I communicated via RS232. Solution was to install the driver. As you supposed in your post.

Felix
0 Kudos
Message 4 of 5
(3,129 Views)
Thank you for your help. It turned out to be the wiring of the cable.
 
The reason it worked in the test environment and not the actuall is a different ports were used. In the test environment a convertor from RS232 was used. The converter required a one wiring which was re-created for a RS485 card which required a different wiring.
Basically there was contention on the liine.