Yes, by "open channel" i mean that there is no signal returning from the gage because it has broken off the structure or the signal to the terminal block has been disconnected somehow.
When I lose data, I'm referring to the data that is acquired on following channels. The behavior appears to be fairly random (which was a troubleshooting nightmare 🙂 ) but usually if a channel is open, then that and those behind it in the scan order do not return any data. It returns a number like 10.357E-6 and I do not pick up any noise or fluctuations. As far as I understand, the multiplexer stops scanning at this point (open channel) because it has nothing to ground to and subsequently "hangs" - not (or randomly) scanning the following channels. I have gone through the bridge configuration and as of yet, have not figured out how to fix this using any hard-wiring techniques. I emailed NI, and was sent a library file to "ground the channels" but it actually resets the Analog input Task I have defined each loop.
I am new to NI-DAQmx so I do not know if this will alter any configurations I already have in MAX for NI-DAQmx. Will it override? I am concerned it will slow my loop time, and also I cannot see the code for this sub.vi so am a little skiddish about using it if i do not know if it can be tailored to my app.