I have a problem with the digital trigger (PFI0). Sometimes it starts without digital signal.

Merci

Jean-Louis SCHRICKE
CTA - Certified TestStand Architect (2008 - 2022)
CTD - Certified TestStand Developer (2004 & 2007)
CLD - Certified LabVIEW Developer (2003 & 2005)

0 Kudos
Message 1 of 3
(2,783 Views)
Not knowing anything about your application, I'll take a guess at this one. I would suspect electrical noise as the most likely culprit. Make sure you are properly grounded and sheilded, and make sure you do not have excesive wire length. You can also try putting a small (pF) capacitor on your input to act as a simple filter on the noise.

Sheilded twisted pair wire can often work miracles in these situations.

Hope this helps.
0 Kudos
Message 2 of 3
(2,783 Views)
Thanks for your answer.

We have found out what was wrong with this digital trigger. It was re-armed to early. The level was still high and there was sometimes a bouncy spike when going to low level.

Best regards.

Jean-Louis SCHRICKE