Hey guys, i am using a labview programm to take I-V measurements with keithley 2604b and when i take the reference (terminals not connected to any device ) i take a linear plot like in the picture. Does that make any sense or the labview Vi that i am using is not so good? 

Thanks a lot !!!!

Reference.JPG

0 Kudos
Message 1 of 3
(1,267 Views)

Hi BuzzG,

 


@BuzzG wrote:

Does that make any sense or the labview Vi that i am using is not so good? 


How can we judge on your LabVIEW VI when you forget to attach that VI?

 

On that image:

You measure ~9e-11A at 10V: that is ~90pA and related to a resistance of ~110GOhm.

What do you expect? And what does Keithley mention in the datasheets and manuals?

Best regards,
GerdW


using LV2016/2019/2021 on Win10/11+cRIO, TestStand2016/2019
0 Kudos
Message 2 of 3
(1,218 Views)

You have a point ! I am attaching the vi .

Thanks alot 

0 Kudos
Message 3 of 3