07-04-2021 11:26 AM
Hey guys, i am using a labview programm to take I-V measurements with keithley 2604b and when i take the reference (terminals not connected to any device ) i take a linear plot like in the picture. Does that make any sense or the labview Vi that i am using is not so good?
Thanks a lot !!!!
07-05-2021 01:46 AM
Hi BuzzG,
@BuzzG wrote:
Does that make any sense or the labview Vi that i am using is not so good?
How can we judge on your LabVIEW VI when you forget to attach that VI?
On that image:
You measure ~9e-11A at 10V: that is ~90pA and related to a resistance of ~110GOhm.
What do you expect? And what does Keithley mention in the datasheets and manuals?
07-05-2021 03:07 AM
You have a point ! I am attaching the vi .
Thanks alot