02-21-2019 08:34 AM
Hi,
I am trying to program Ismatec Reglo Digital pumps to control it through my pc. I downloaded the LABVIEW driver from Ismatec website
but when i open the any VI file and run it, it gives me error out 7001.
I do not know exactly what is the reason behind this error because i tried to change the cable but i get the same problem. Any suggestions would be very much appreciated.
Thanks.
02-21-2019 08:58 AM
With new equipment I usually try using the manufacturer software first to make sure everything is working properly. Then try using the LabVIEW drivers with the settings that worked.
I have never used one of those pumps, so I have no idea how well their drivers are written. But some drivers are written better then others.
02-21-2019 09:02 AM
@_Bryan wrote:
With new equipment I usually try using the manufacturer software first to make sure everything is working properly. Then try using the LabVIEW drivers with the settings that worked.
I have never used one of those pumps, so I have no idea how well their drivers are written. But some drivers are written better then others.
Thanks Bryan for your quick reply.
I already used the software written by Ismatec but it is not working.
Thanks you.
02-21-2019 09:10 AM
@Darawish wrote:
I already used the software written by Ismatec but it is not working.
If it does not work with the supplied software it will not magically work with LabVIEW. Make sure the communication settings are the same on the unit and in the software and then double check the wiring. You could also contact the pumps support department to get it working with their software first.